Ukuhlolwa kokusebenza phakathi kwe-SF-8200 ne-Stago Compact Max3


Umbhali: Olandelayo   

微信图片_20211012132116

I-artical yashicilelwa ku-Clin.Lab.by Oguzhan Zengi, Suat H. Kucuk.

Yini i-Clin.Lab.?

I-Clinical Laboratory iyiphephabhuku lamazwe ngamazwe elibuyekezwe ontanga ngokugcwele elihlanganisa zonke izici zokwelapha zaselabhorethri kanye nokwelapha ngokumpontshelwa igazi.Ngaphezu kwezihloko zemithi yokumpontshelwa igazi I-Clinical Laboratory imelela ukuhanjiswa okuphathelene nokufakelwa kwezicubu kanye nokwelashwa kwe-hematopoietic, amaselula kanye nezakhi zofuzo.Ijenali ishicilela izindatshana zoqobo, izindatshana zokubukeza, amaphosta, imibiko emifushane, izifundo zezindaba kanye nezincwadi eziya kumhleli obhekene 1) nesizinda sesayensi, ukuqaliswa kanye nokubaluleka kokuxilonga kwezindlela zaselabhorethri ezisetshenziswa ezibhedlela, ezindaweni zokugcina igazi nasemahhovisi odokotela kanye no-2) izici zesayensi, zokuphatha nezomtholampilo zomuthi wokumpontshelwa kanye 3) ngaphezu kwezihloko zomuthi wokumpontshelwa Ilabhorethri Yomtholampilo imelela ukuhanjiswa okuphathelene nokufakelwa kwezicubu kanye nokwelashwa kwe-hematopoietic, amaselula kanye nofuzo.

 

elebhu yomtholampilo

Babehlose ukwenza ucwaningo lokuqhathanisa ukusebenza kohlaziyo phakathi kwe-Succeeder SF-8200 ne-Stago Compact Max3 ngoba

abahlaziyi be-coagulation abazenzakalelayo sebephenduke enye yezingxenye ezibaluleke kakhulu zamalabhorethri emtholampilo.

Izindlela: Ukuhlolwa kwe-coagulation okujwayelekile kwahlolwa, okuyizona ezi-odiwe kakhulu kumalabhorethri afana ne-PT, APTT, ne-fibrinogen.

Imiphumela: Ama-coefficients okuhlukahluka ahlolwe ekuhlaziyweni kokunemba kwe-intra kanye ne-inter-assay ayengaphansi kuka-5% ngokumele amapharamitha ahloliwe.Ukuqhathanisa phakathi komhlaziyi kubonise imiphumela emihle.Imiphumela etholwe yi-SF-8200 ibonise ukuqhathaniswa okuphezulu kakhulukazi kubahlaziyi bereferensi abasetshenzisiwe, namakhoyefiyenti okuxhumana asuka ku-0.953 kuya ku-0.976.Esimisweni sethu saselabhorethri esijwayelekile, i-SF-8200 ifinyelele izinga lokuphuma kwesampula lama-360 ngehora.Alikho ithonya elikhulu ekuhlolweni elitholakele lamazinga aphakeme e-hemoglobin yamahhala, i-bilirubin, noma i-triglycerides.

Iziphetho: Ekuphetheni, i-SF-8200 yayiyi-analyzer enembile, enembile, futhi enokwethenjelwa ekuhloleni okujwayelekile.Ngokocwaningo lwethu, imiphumela ibonise ukusebenza okuhle kakhulu kwezobuchwepheshe nokuhlaziya.